CA75 DPD analyzer module for total and free chlorine in drinking water, wastewater, process water

CA75 DPD analyzer module for total and free chlorine

Colorimetric analyzer module for online monitoring of drinking water, wastewater and process water

CA75
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Free chlorine

  • Application

    Monitoring disinfection in drinking water treatment
    Measuring chlorine levels in secondary water supply systems
    Monitoring process and utility water in the food and beverage industry
    Chlorine monitoring in industrial water circuits
    Analyzing chlorine levels in water reuse applications

  • Characteristic

    Colorimetric measurement of free chlorine or total chlorine

  • Measuring range

    0 to 10 mg/l free chlorine or 0 to 10 mg/l total chlorine
    LOD: 0.018 mg/l

  • Measuring method

    DPD (N,N-Diethyl-p-phenylenediamine) method in accordance with US EPA 40 CFR 141.74, US EPA 40 CFR 136.3, US EPA 334.0, ISO 7393-2 and 4500-CL-G

  • Design

    Compact DPD analyzer module (to be connected to Liquiline CM44x)

  • Dimension

    449 x 340 x 199 mm

  • Process temperature

    5 to 40 °C (41 to 104 °F)

  • Process pressure

    0.3 bar (4.4 psi) to 5.2 bar (75.4 psi), up to the point of sample preparation.

  • Sample flow rate

    60 to 200 ml/min

  • Connection

    Fixed cable with Memosens-Technology

  • Supply voltage

    via Liquiline CM44x transmitter

  • Consumables

    Reagents and standard solutions CY75
    Cleaner CY800
    Maintenance kit CAV750